• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL Back to Search Results
Model Number MCENDO550R
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that the scope was very dark.It is unknown when this event occurred, whether the product was changed out, or if there was any effect on the patient or results of the surgery.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.
 
Event or Problem Description
The event occurred during setup.The product was changed out.There was no delay and the procedure was completed.Successfully with no patient effect.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on december 17, 2018.(b)(4).The sample was not returned for evaluation; however, past product complaints were reviewed for similar issues and possible causes would be the product was either dropped or inadvertently struck by an object and excessive force was exerted on the endoscope shaft; this could deflect the shaft and lead to the alteration of the internal lens system.The labeling for the device stresses the importance of how fragile the device is and instructs the user to handle the product with extreme care to prevent damage to the device.The most likely root cause is improper handling caused the internal lens to break.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURBISHED ENDOSCOPE 5.5M
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8172299
Report Number1124841-2018-00329
Device Sequence Number16518524
Product Code GCJ
UDI-Device Identifier00699753023222
UDI-Public(01)00699753023222
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K092789
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMCENDO550R
Device Catalogue NumberN/A
Device Lot Number781429
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/28/2018
Supplement Date Received by Manufacturer01/05/2019
Initial Report FDA Received Date12/17/2018
Supplement Report FDA Received Date01/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-