U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G31519
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
510 (k) number; k083330.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Lab evaluation the device related to this occurrence underwent a laboratory evaluation on 07 dec 2018.A kink was observed proximal to the sheath extender.Document review prior to distribution, all echo-25 devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-25 of lot number c1505525 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1505525.Ifu review the notes section of the instructions for use, ifu0101-0 which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use (ifu0101-0).Root cause review the failure of proximal kink was observed in the laboratory.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible cause could be attributed to damage caused during transit and/or storage.Summary complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
"needle was bent coming out of the packaging" will be returned.Not used.Lot confirmed."as per complaint form": nurse opened package and noticed the needle was bent/kinked.Tried to bend it back however doctor refuses to use the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8172785
Report Number3001845648-2018-00600
Device Sequence Number16495437
Product Code FCG
UDI-Device Identifier00827002315198
UDI-Public(01)00827002315198(17)210601(10)C1505525
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K083330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2021
Device Model NumberG31519
Device Catalogue NumberECHO-25
Device Lot NumberC1505525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/17/2018
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2018
Initial Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-