Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation evaluation.A review of the dimensional verification, complaint history, device history record, manufactures instructions, quality control, trends, and a functional test & visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the dilator was inserted partially into the device.There was a kink in the introducer tubing at 2.4cm from the cap.The device has biological matter throughout.The tubing was occluded with hemostats and water was injected into the side arm to test for leakage.The device leaked around the silicone disc.The connector cap was disassembled from the check flo body and the cap-to-bottom distance is 1.388¿.There was glue on the threads of the fitting, and no glue in the cap.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should also be noted that the only other complaints associated with this lot number as all from the same facility and for the same failure mode.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Measures are being conducted to address this failure mode.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|