TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Codes Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Dizziness (2194)
Date of Event 11/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-30-user applied blood to strip before inserting strip into meter test strip udi#: (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product notification letter to customer to contact customer care because customer address was not provided.
 
Event or Problem Description
Consumer reported complaint for e-3 error accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptom of feeling lightheaded.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product storage location is undisclosed.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test result of 100mg/dl.Test strip lot manufacturer's expiration date is 05/31/2019, and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.Asked customer how she was feeling and customer states she feels lightheaded.Asked if customer was ok to continue with the product troubleshooting and customer states if she needs to go to the hospital she will, but at this time, she states she is ok enough to continue with the call.Customer is concerned of e-3 once the strip is inserted.Customer states she inserts the strip into the meter with the blood sample.While troubleshooting, customer ran a blood test and obtained a reading of 100mg/dl.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8174491
Report Number1000113657-2018-01258
Device Sequence Number1747808
Product Code NBW
UDI-Device Identifier00021292007447
UDI-Public(01)00021292007447
Combination Product (Y/N)N
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2707
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2018
Initial Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No