Catalog Number 115306
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment an external fluid leak occurred at the top of the venous line near the blood port of a prismaflex m100 set.An alarm was triggered.There was patient involvement but no patient impact and no medical intervention.No additional information available.
 
Manufacturer Narrative
Additional information: the actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided pictures shows that the return line was not fully seated in the connector at the top of the filter.The return line was ungluing which caused the reported leak.The reported condition was confirmed.The cause was determined to be a manufacturing defect and an operator error.The operator of the device has been retrained.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M100 SET
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
na
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8174613
MDR Text Key131793283
Report Number8010182-2018-00175
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K041005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2020
Device Catalogue Number115306
Device Lot Number18D2709
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/26/2018
Initial Date FDA Received12/18/2018
Supplement Dates Manufacturer Received01/16/2019
Supplement Dates FDA Received01/25/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-