Reported event: an event regarding audible noise (squeaking) involving a ceramic head was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: a review of the provided medical records and/or x-rays by a clinical consultant indicated- this requires x-rays for evaluation.There are only radiological reports available, no actual x-rays, to describe the arthroplasty in ¿anatomic conditions¿.Unfortunately however, there are few radiologists with a full understanding of the intricate and sometimes subtle biomechanics issues associated with suboptimal component position in arthroplasties.So, real x-rays are required.The fact of an impingement problem is further supported by the presence of a noise problem since time of implantation to suggest indeed a biomechanical type of derangement in the arthroplasty.So, component malposition is suspect, frequently the cup, unless proof of contrary through presence of good quality x-rays including preferably a true lateral x-ray for measurement of cup anteversion.With the current info, no exact cause of the problem can be defined although component impingement of some type remains highly suspect.Further evaluation of the case is suggested with ct-scan to determine exact component positions if the problem cannot be solved with standard xrays.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the event could not be confirmed and the exact cause of the event could not be determined because insufficient information was provided.A review of the provided medical records and/or x-rays by a clinical consultant indicated- this requires x-rays for evaluation.There are only radiological reports available, no actual x-rays, to describe the arthroplasty in ¿anatomic conditions¿.Unfortunately however, there are few radiologists with a full understanding of the intricate and sometimes subtle biomechanics issues associated with suboptimal component position in arthroplasties.So, real x-rays are required.The fact of an impingement problem is further supported by the presence of a noise problem since time of implantation to suggest indeed a biomechanical type of derangement in the arthroplasty.So, component malposition is suspect, frequently the cup, unless proof of contrary through presence of good quality x-rays including preferably a true lateral x-ray for measurement of cup anteversion.With the current info, no exact cause of the problem can be defined although component impingement of some type remains highly suspect.Further evaluation of the case is suggested with ct-scan to determine exact component positions if the problem cannot be solved with standard xrays.If further information becomes available or the product is returned, this investigation will be re-opened.Device not returned.
|
|
It was reported that patient has experienced audible noise in their right hip since implantation (noise reported as being squeaky).At first, the noise was intermittent, but has increased in frequency and has become more pronounced in the last 6 months.The surgeon wants to know if similar reports have been received for these devices as the surgeon cannot determine if the squeaking arises from the head/trunion interface or the shell/liner interface.The patient has not yet been revised, surgeon wants to wait for the aforementioned results before making a decision.
|