Model Number 10604
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in the posterolateral (pl) branch of the right coronary artery (rca).A 2.75x24mm synergy drug-eluting stent was advanced for treatment.However, the device was unable to pass the moderate to severe calcium in the mid rca.It was felt that the struts were lifted when trying to advance the device in the mid area of the rca.The device was removed and took another of the same device.The new stent was able to pass the mid vessel and was successfully placed in the pl of the rca and completed the procedure.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy ous mr 2.75 x 24 mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and the inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No issues were identified during the product analysis.Device is a combination product.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in the posterolateral (pl) branch of the right coronary artery (rca).A 2.75x24mm synergy drug-eluting stent was advanced for treatment.However, the device was unable to pass the moderate to severe calcium in the mid rca.It was felt that the struts were lifted when trying to advance the device in the mid area of the rca.The device was removed and took another of the same device.The new stent was able to pass the mid vessel and was successfully placed in the pl of the rca and completed the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8174805
Report Number2134265-2018-64478
Device Sequence Number1626337
Product Code NIQ
UDI-Device Identifier08714729840220
UDI-Public08714729840220
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2020
Device Model Number10604
Device Catalogue Number10604
Device Lot Number0022519045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Initial Date Received by Manufacturer 12/11/2018
Supplement Date Received by Manufacturer01/30/2019
Initial Report FDA Received Date12/18/2018
Supplement Report FDA Received Date02/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-