MEDTRONIC PUERTO RICO OPERATIONS CO. VISIA AF XT VR ICD; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DVAB1D4
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 12/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented to the emergency department after receiving therapy from their implantable cardioverter defibrillator (icd).A remote transmission was sent.Information received on the remote transmission was reviewed by qualified personnel.It was determined that the device had misdetected atrial fibrillation with rapid ventricular response (af w/rvr) as a ventricular fibrillation (vf) episode, resulting in the patient receiving an inappropriate therapy.Follow up was conducted with the patients listed clinic.The healthcare professional (hcp) was able to verify that the patient has not been seen in-office yet and therefore, no reprogramming has been completed at this time.The device remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIA AF XT VR ICD
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key8174886
Report Number3004209178-2018-27912
Device Sequence Number19493169
Product Code LWS
UDI-Device Identifier00643169720626
UDI-Public00643169720626
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/28/2019
Device Model NumberDVAB1D4
Device Catalogue NumberDVAB1D4
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2018
Initial Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M62, LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age89 YR
-
-