STRYKER NEUROVASCULAR-CALIF AXS INFINITY LS PLUS LONG SHEATH (90CM); CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number INC-11196-90
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that the catheter shaft was found to be broken when the package was opened during preparation for the procedure.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.Visual examination of the returned device revealed that the catheter distal shaft was kinked on its proximal section just at strain relief distal end.The catheter was found to be separated as well.No other anomalies were noted.Functional testing could not be performed due to the damaged condition of the returned catheter.It is possible that the catheter was kinked first and then fractured due to excessive lifting force applied to the hub/strain relief during removal from its packaging.Therefore based on all of the information available, the catheter broke outside the patient due to handling.Further review of the reported event indicated that the reported catheter break occurred outside the patient during preparation for the procedure and procedure was completed using a new catheter.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Event or Problem Description
It was reported that the catheter shaft was found to be broken when the package was opened during preparation for the procedure.There were no reported clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS INFINITY LS PLUS LONG SHEATH (90CM)
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR-CALIF
47900 bayside parkway
fremont CA 94538
MDR Report Key8175032
Report Number3008853977-2018-00078
Device Sequence Number2178550
Product Code DQY
UDI-Device Identifier07613327459203
UDI-Public07613327459203
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K152876
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2021
Device Catalogue NumberINC-11196-90
Device Lot Number13203-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Initial Date Received by Manufacturer 11/27/2018
Supplement Date Received by Manufacturer01/15/2019
Initial Report FDA Received Date12/18/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-