• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VERSALOK THREADER TAB *EA; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VERSALOK THREADER TAB *EA; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 210800
Device Problem Peeled/Delaminated (1454)
Patient Problem Not Applicable (3189)
Event Date 02/13/2018
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).Investigation summary: the complaint device is not available for physical evaluation, hence, the complaint cannot be confirmed.Given the information provided we cannot discern a definitive root cause for the reported failure.A device history record (dhr) review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliate in (b)(6) that during rotator cuff repair surgical procedure, it was observed that there were metallic matter floating arthroscopically.It was reported that there was no breakage nor anomaly found in the device.It was reported that the device was brand new and on its first use when the issue occurred.There was a 10-minute-delay in surgery and no harm to the patient.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8175208
MDR Text Key131213445
Report Number1221934-2018-55719
Device Sequence Number1
Product Code HWC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063478
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Catalogue Number210800
Device Lot NumberL621152
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/14/2018
Initial Date FDA Received12/18/2018
Date Device Manufactured11/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-