| Catalog Number |
JJ41G |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During the procedure, the or staff opened the sterile suture packet and a different type of suture and needle was inside.Suture from a different lot were used to complete the procedure.There were no patient consequences reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 01/15/2019.Investigation summary: it was reported assembly incorrect component.An empty opened foil of product code jj41, lot # mgz466 and a labeled winding former with eight parked needle/suture combinations from slot # 1 to 8 of product code j774 were returned for analysis.During the visual inspection of sample, a mismatched between the product code jj41 of foil and the product code j774 of paper lid of winding former were observed.Also, the samples into of winding former were examined and belong to product code j774.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the sample received, the assignable cause of assembly incorrect component is mixed product code.During the visual inspection of six unopened samples, no defects were found on the package.The samples were opened and it was observed a mismatched in one sample between the product code jj41 of foil and the product code j774 of paper lid of winding former were observed.Also, the samples into of winding former were examined and belong to product code j774.The five samples were examined and no incorrect component were observed.
|
| |
|
Search Alerts/Recalls
|