• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURN ART INS BCS STD 7-8 LT13; PRSTHSIS,KNEE,PATELLOFMOROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/MTAL/POLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURN ART INS BCS STD 7-8 LT13; PRSTHSIS,KNEE,PATELLOFMOROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/MTAL/POLYMR Back to Search Results
Catalog Number 74023284
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/21/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed.Patient presented with metallosis, joint was inflamed, soft tissue had a black colour.The surgeon revised the knee, took out the implants and replaced them.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.The clinical / medical investigation concluded that, since no supporting clinical documentation has been provided for review, the reported findings cannot be confirmed.Without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the source cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the reported symptoms, the intra-op findings with suspected titanium allergy, and the revision surgery with an expected post-operative rehabilitative phase cannot be concluded.Without the actual product involved, our investigation cannot proceed.If the devices or new information is received in the future, these complaints can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURN ART INS BCS STD 7-8 LT13
Common Device Name
PRSTHSIS,KNEE,PATELLOFMOROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/MTAL/POLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8175536
Report Number1020279-2018-02824
Device Sequence Number13890169
Product Code JWH
UDI-Device Identifier03596010555373
UDI-Public03596010555373
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/14/2017
Device Catalogue Number74023284
Device Lot Number07DM11500
Initial Date Received by Manufacturer 11/22/2018
Supplement Date Received by Manufacturer11/22/2018
Initial Report FDA Received Date12/18/2018
Supplement Report FDA Received Date05/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-