SMITH & NEPHEW, INC. JOURNEY TIBIA BASE NP LT SZ 7; PRSTHSIS,KNEE,PATELLOFMOROTIBIAL,SMI-CNSTRAIND,CMNTD,POLYMR/MTAL/POLYMR
|
Back to Search Results |
|
| Catalog Number |
74022227 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
11/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed.Patient presented with metallosis, joint was inflamed, soft tissue had a black colour.The surgeon revised the knee, took out the implants and replaced them.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were not returned.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.The clinical / medical investigation concluded that, since no supporting clinical documentation has been provided for review, the reported findings cannot be confirmed.Without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the source cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the reported symptoms, the intra-op findings with suspected titanium allergy, and the revision surgery with an expected post-operative rehabilitative phase cannot be concluded.Without the actual product involved, our investigation cannot proceed.If the devices or new information is received in the future, these complaints can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|