| Catalog Number |
6725137105 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is available for return.A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty that two precision blades broke at the contact point of the precision saw.It was further reported that there was a short delay to get a new blade.It was also reported that there were no adverse consequences and that the procedure was completed successfully.This report is for the second blade that broke.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 & h6 investigation results indicate that the most probable root cause of the cartridge break is as a result of issues with the handpiece (b)(6).The td ost shaft broke between the link and the bottom bearing in the associated handpiece.The quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty that two precision blades broke at the contact point of the precision saw.It was further reported that there was a short delay to get a new blade.It was also reported that there were no adverse consequences and that the procedure was completed successfully.This report is for the second blade that broke.
|
| |
|