| Catalog Number |
0279351250 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe would not turn off.
|
| |
|
Event or Problem Description
|
|
It was reported that the probe would not turn off.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: per customer the probe would not turn off.The failure identified in the investigation is consistent with the complaint record.The probable root cause/s could be the detachment of the suction tube to the suction shaft causing a leak internally, which creates a short and a possibility would result in continuous activation.Inadequate gluing operation on the suction tube and suction adapter.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
|
| |
|