• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7130
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/06/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 99% stenosed target lesion was located in the right coronary artery.A 6fr jr4 non-bsc guide catheter was used and a 180cm non-bsc guide wire was advanced to cross the lesion.A 1.50mm x 8mm emerge balloon catheter was advanced for dilatation; however, it was noted that the shaft was fractured.The procedure was completed with a different device.No patient complications were reported and the patient's status was fine.
 
Event or Problem Description
It was reported that shaft break occurred.The 99% stenosed target lesion was located in the right coronary artery.A 6fr jr4 non-bsc guide catheter was used and a 180cm non-bsc guide wire was advanced to cross the lesion.A 1.50mm x 8mm emerge balloon catheter was advanced for dilatation; however, it was noted that the shaft was fractured.The procedure was completed with a different device.No patient complications were reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 71.5cm from the hub.Microscopic examination revealed no additional damages.There is blood present in the guide wire lumen and the balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8176917
Report Number2134265-2018-64365
Device Sequence Number19519293
Product Code LOX
UDI-Device Identifier08714729806493
UDI-Public08714729806493
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121196
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2020
Device Model Number7130
Device Catalogue Number7130
Device Lot Number0021506017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Initial Date Received by Manufacturer 12/06/2018
Supplement Date Received by Manufacturer01/14/2019
Initial Report FDA Received Date12/18/2018
Supplement Report FDA Received Date02/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-