• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 7N8376K
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a one-link y-type standard bore catheter extension set would not flush.This occurred during an unspecified process step.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The device was received in a zip lock bag inside the opened blister.Visual inspection did not identify any abnormalities that could have contributed to the reported condition, all components were correctly placed according to product specifications.Functional tests (pressure test, clear passage and block test) were performed and the results were satisfactory; the sample flowed normally.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina, san cristobal
MDR Report Key8177030
Report Number1416980-2018-07881
Device Sequence Number2025575
Product Code FPA
UDI-Device Identifier00085412599601
UDI-Public(01)00085412599601
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K132734
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/26/2022
Device Catalogue Number7N8376K
Device Lot NumberDR17E25098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Initial Date Received by Manufacturer 11/27/2018
Supplement Date Received by Manufacturer01/09/2019
Initial Report FDA Received Date12/18/2018
Supplement Report FDA Received Date01/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-