PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 12 DEFIBRILLATOR/MONITOR SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 99400-000822
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer, a biomedical engineer, evaluated the device and verified the reported issue.(b)(6) provided the biomed with technical assistance.The customer determined that the cause of the reported issue was due to the damaged therapy connector on the device.Physio provided the customer with the therapy connector repair kit part number.The device was not returned to (b)(6) for an evaluation.
 
Event or Problem Description
The customer, a biomedical engineer, contacted (b)(6) to report that the therapy connector on their device is damaged and the therapy cable is difficult to attach to the device.This issue has the potential to either delay, or prevent, defibrillation from being delivered.There was no patient use associated with the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 12 DEFIBRILLATOR/MONITOR SERIES
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key8177054
Report Number0003015876-2018-02145
Device Sequence Number16508806
Product Code MKJ
UDI-Device Identifier00721902152962
UDI-Public00721902152962
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K102972
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number99400-000822
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2018
Initial Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-