STRYKER ENDOSCOPY-SAN JOSE PKG, STRYKERFLOW ELECTROCAUTERY PROBE, J TIP, 5MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0250070442
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.The reported failure mode will be monitored for future reoccurrence.Alleged failure: sheath melted.Probable root cause: material/design error, manufacturing/assembly error, excessive user force, severe shipping conditions, disposable tip rubbing against product, user error.The device manufacture date is not known.
 
Event or Problem Description
It was reported that the sheath broke during a procedure and cannot be confirmed that all pieces were retrieved.
 
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.The reported failure mode will be monitored for future reoccurrence.Alleged failure: sheath melted.Probable root cause: 1.Material/design error 2.Manufacturing/assembly/ service error 3.Excessive user force 4.Severe shipping conditions 5.Disposable tip rubbing against product 6.User error the device manufacturer date is not known.
 
Event or Problem Description
It was reported that the sheath broke during a procedure and cannot be confirmed that all pieces were retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, STRYKERFLOW ELECTROCAUTERY PROBE, J TIP, 5MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8177055
Report Number0002936485-2018-00721
Device Sequence Number11096053
Product Code GCJ
UDI-Device Identifier07613327051889
UDI-Public07613327051889
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K934094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0250070442
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/26/2018
Supplement Date Received by Manufacturer11/26/2018
Initial Report FDA Received Date12/18/2018
Supplement Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
Patient Weight80
-
-