| Model Number |
Z9002 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as the product was not implanted.If explanted, give date: not applicable, as the product was not implanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that when a z9002 intraocular lens was received, the product had a damaged haptic.It was broken when received but the boxes where not damaged.There was no patient contact.No additional information was received.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received from the investigation confirmed the complaint product was handled.As a result, this event has been re-assessed as not reportable.Device available for evaluation: yes, returned to manufacturer on: 1/16/2019, device returned to manufacturer: yes.Device evaluation: the product was received on 01/16/2019 in the original folding carton.Two lenses were received.One lens s cut in pieces.The other is with a detached haptic.Traces of what appears to be viscoelastics and /or balanced salt solution are observed on the lenses surface; indicating the lens was handled to be prepared for insertion.The haptics are torn or broken as described on the initial report.The reported issue was verified.A product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional investigation request for this production order number has been received.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|