| Model Number |
DLA19 |
| Medical Device Problem Codes |
Calcified (1077); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Aortic Insufficiency (1715); Aortic Valve Stenosis (1717); No Information (3190)
|
| Date of Event |
11/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device returned to the manufacturer on (b)(4) 2018 for analysis.The gross examination conducted on (b)(4) 2018 confirmed that the returned valve appeared heavily deteriorated with traces of organic deposition.Residues of stitches were also present.Further investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
In 2013 (exact date unknown), the patient received a mitroflow in the aortic position.On (b)(6) 2018, the patient was re-operated to have the mitroflow explanted and replaced with a st.Jude 19mm.Two other concomitant procedures were performed on the same day: mitral valve replacement with a 27mm st.Jude and tricuspid repair with an edwards ring 30mm.No other information is available to date.
|
| |
|
Additional Manufacturer Narrative
|
|
The visual examination, xray analysis, and histopathological exam were performed on this mitroflow valve sn (b)(6), size 19.The results showed that the leaflets calcification, detected with visual, x-ray and histological analyses, caused stiffening and led to progressive valve stenosis.Aortic insufficiency likely resulted from leaflets tears and inadequate leaflets coaptation, caused by calcification of the leaflets.There was no evidence of endocarditis in the returned valve, confirming the information reported on this event.Furthermore, the manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # dla19, s/n # (b)(6), as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova canada corp.The results confirmed that this valve satisfied all material, visual, and performance standards required for a dla19 mitroflow aortic pericardial heart valve at the time of manufacture and release.Based on the performed analysis and the information received from the site, it is not possible to determine the exact root cause at this time.However, as reported in the scientific literature, structural dysfunction is the major cause of failure of bioprosthetic heart valves and the principal underlying pathologic process is cuspal calcification.Calcification can also cause stenosis due to cuspal stiffening.Calcific deposits are usually localized to cuspal tissue (intrinsic calcification).
|
| |
|
Event or Problem Description
|
|
In 2013 (exact date unknown), the patient received a mitroflow in the aortic position.On (b)(6) 2018, the patient was re-operated to have the mitroflow explanted due to severe stenosis and moderate insufficiency, and replaced with a st.Jude 19mm.Two other concomitant procedures were performed on the same day: mitral valve replacement with a 27mm st.Jude and tricuspid repair with an edwards ring 30mm.It is reported that the patient had a febrile episode in august 2018 and, after about a month, she had dyspnea on moderate exertion.On (b)(6) 2018 she was hospitalized, and a blood culture test was performed which came out negative.At the echocardiogram, there was dysfunction of the aortic bioprosthesis which resulted in severe stenosis.Since endocarditis was suspected, triple antibiotic therapy was started with vancocin, gentamicin, and rifampicin.After 13 days from the beginning of antibiotic therapy, she was transferred to the european hospital for surgery.The tt echocardiogram detected: fibrocalcific degeneration of semilunar determining severe stenosis and insufficiency of moderate degree; severe stenosis and moderate insufficiency of the mitral valve (calcification of the flaps and annulus); moderate insufficiency of the tricuspid; no pericardium effusion.Intraoperatively, the valve showed marked calcifications without evidence of perivalvular leaks.No signs of vegetations, perforations of the flaps, nor evident lesions from previous endocarditis were observed.The postoperative course was prolonged due to an initial phase of heart failure with significant water retention, and also due to the need to carry out a prolonged cycle of antibiotic therapy adequate to the vegetation suspected pre-operatively (then averted by the intraoperative findings).The patient is currently in excellent post-operative conditions.
|
| |
|