| Model Number |
10007 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Ischemia (1942); Vascular System (Circulation), Impaired (2572)
|
| Date of Event |
12/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: visual inspection shows a piece of the broke device was returned installed on a connector.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of a dlp perfusion adapter when the patient (on full va-ecmo support) was been brought from intensive care to the operating room the connector broke.The break partially stopped blood circulation to the patient for approximately 2 minutes.The adapter had been in place for 4 days.There was no leak as air was aspirated through the broken connector.It is unknown if the device was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion; complaint is confirmed for a broken perfusion adapter.It is unknown what may have caused this occurrence, however there is the possibility that this was caused by excessive force being applied during use.Correction: patient code updated if information is provided in the future, a supplemental report will be issued.
|
| |
|