| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a power (battery life w/ damage) issue.There was no allegation of an adverse event associated with this complaint.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 17-jan-2019 with the following findings: investigation confirmed frequent low battery warnings in the alarm history.The black box data revealed low battery voltage immediately following battery change.The pump¿s electrical current draws were found to be within specifications.Investigation did not duplicate the complaint.Unrelated to the original complaint, the battery compartment was noted to be cracked during the investigation.
|
| |
|