• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9438-06
Device Problem No Device Output (1435)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that the display device showed a transmitter failed error on (b)(6) 2018.Data was returned and evaluated.The reported event of a transmitter failed error was confirmed via data.The probable cause was determined to be a low transmitter battery that led to a transmitter failure.No additional patient or event information is available.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.An external visual inspection was performed and passed.Voltage testing was performed, and the test failed with 0 voltage.A review of the share logs was performed and a transmitter failed error was found within the investigation window.The probable cause was determined to be a low transmitter battery.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key8178857
MDR Text Key131025306
Report Number3004753838-2018-158576
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9438-06
Device Catalogue NumberSTT-GF-005
Device Lot Number7243747
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/21/2018
Initial Date FDA Received12/19/2018
Supplement Dates Manufacturer Received07/24/2019
Supplement Dates FDA Received07/26/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/27/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age10 YR
Patient Weight45
-
-