| Model Number |
10605 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and removal difficulties occurred.The target lesion was located in the left anterior descending artery.A 2.75 x 28mm synergy ii drug-eluting stent was advanced to treat the lesion.However, when negative pressure was applied with the indeflator, it was noted in the hub of the delivery catheter as well as in the indeflator tubing, that the balloon ruptured.The physician was unable to pull the stent back into the guide, so the wire, stent delivery catheter, and guide were removed.Another 2.75x28mm synergy stent was deployed and the procedure was completed.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr.: synergy ii us mr 2.75 x 28 mm stent delivery system was returned for analysis.A visual examination of the stent found that the proximal end of stent was damaged with stent struts bunched distally and the distal end of stent was also damaged with stent struts pulled distally.The undamaged stent od (outer diameter) was measured within maximum crimped stent profile measurement.Stent damage most likely occurred during withdrawal attempts.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.The device was returned loaded in guide catheter with a 0.070'" guidewire in the wire lumen of the device.The hypotube shaft was not visible.As part of analysis an attempt was made to inflate the device to rbp (rated burst pressure) however the device-maintained pressure but failed to inflate.The device was soaked in the water-bath to soften hardened medium that could have blocked the wire lumen.After attempting to inflate again it was observed that the device was leaking from the guide catheter.The device was cut 15 mm distal from the distal end of the strain relief to remove the device through the tip of the guide catheter and a visual and tactile examination of the hypotube found a hypotube break at 670 mm distal from the distal end of the strain relief.This type of damage is consistent with excessive force that could have been applied on the delivery system during advancing attempts.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues with the extrusion shaft.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and removal difficulties occurred.The target lesion was located in the left anterior descending artery.A 2.75 x 28mm synergy ii drug-eluting stent was advanced to treat the lesion.However, when negative pressure was applied with the indeflator, it was noted in the hub of the delivery catheter as well as in the indeflator tubing, that the balloon ruptured.The physician was unable to pull the stent back into the guide, so the wire, stent delivery catheter, and guide were removed.Another 2.75x28mm synergy stent was deployed and the procedure was completed.No patient complications were reported and the patient's status was stable.
|
| |
|