• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. LAP-BAND SYSTEM 10CM; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. LAP-BAND SYSTEM 10CM; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number B-2220
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Dysphagia/ Odynophagia (1815); Regurgitation (2259)
Date of Event 10/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Unknown taper type.Device labeling addresses the reported event as follows: adverse events it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Warnings: patients should be advised that the lap-band ap® system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event or Problem Description
Reported as: a patient with the lap-band system was reported to have "gerd/reflux and dysphagia." the device was removed.
 
Additional Manufacturer Narrative
Taper ii.Supplement # 1 - medwatch sent to the fda on 25/jan/2019.Additional information: device evaluation summary: the device was returned to apollo for analysis, and visual examination noted the lap-band with access port i, taper ii was received.The band tubing was separated approximately 1 inch from the port tubing ss connector.Needle marks were observed on the port septum.White and brown particles were noted on the inner surface of the band ring.The band was noted to be separated in half near the band buckle the separated portion of the band with the buckle noted to be still strapped to the band belt.An air leak test was performed and leakage was observed from the band ring and shell.A fill inspection test was performed and no blockage was observed.Under microscopic analysis both ends of the separated band ring and shell and both ends of the band tubing were noted to have striated edges consistent with damage from a surgical tool.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM 10CM
Common Device Name
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key8179145
Report Number3006722112-2018-00335
Device Sequence Number1995069
Product Code LTI
UDI-Device Identifier10811955020169
UDI-Public10811955020169
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P000008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/03/2007
Device Model NumberB-2220
Device Catalogue NumberB-2220
Device Lot Number1064255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 11/19/2018
Supplement Date Received by Manufacturer02/26/2018
Initial Report FDA Received Date12/19/2018
Supplement Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient Weight149
-
-