• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24690
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/03/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft fracture occurred.A 3.0mmx30mmx135cm (4f) sterling balloon catheter was selected for dilatation.However, the device was found fractured after unpacking before usage.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned device consisted of a sterling mr balloon catheter.The balloon was tightly folded.The tip, balloon, markerbands, proximal weld, and inner/outer shafts were microscopically and visually inspected.Inspection revealed numerous kinks in the hypotube.Inspection of the remainder of the device presented no additional damage or irregularities.The hypotube kinks are attributed to handling of the device.
 
Event or Problem Description
It was reported that shaft fracture occurred.A 3.0mmx30mmx135cm (4f) sterling balloon catheter was selected for dilatation.However, the device was found fractured after unpacking before usage.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8179163
Report Number2134265-2018-64508
Device Sequence Number19291066
Product Code DQY
UDI-Device Identifier08714729756873
UDI-Public08714729756873
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K141150
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/12/2021
Device Model Number24690
Device Catalogue Number24690
Device Lot Number0022508220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Initial Date Received by Manufacturer 12/07/2018
Supplement Date Received by Manufacturer01/30/2019
Initial Report FDA Received Date12/19/2018
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight80
-
-