MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT0408
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pet lock was intact to the proximal end of the pusher assembly.The pusher assembly mid-joint was proximal to the introducer sheath friction lock.The embolization coil was intact with the pusher assembly and the coil was undamaged.Evaluation of the returned smart coil revealed a functional device.The introducer sheath friction lock inner diameter (id) is smaller than the rest of the introducer sheath and allows it to seat on the pusher assembly mid-joint.If the mid-joint is retracted proximal to the introducer sheath friction lock, resistance may be experienced when attempting to advance the device out of its introducer sheath.The resistance experienced may have contributed to the inability to advance the device through its introducer sheath.During functional testing, the introducer sheath was reseated properly on the pusher assembly mid-joint, and was able to advance through a demonstration microcatheter without issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, the physician successfully placed three smart coils using a non-penumbra microcatheter.The physician then was unable to advance a smart coil within its introducer sheath.Therefore, the smart coil was removed.The procedure was completed using a new smart coil and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8179453
Report Number3005168196-2018-02469
Device Sequence Number19311490
Product Code HCG
UDI-Device Identifier00814548015392
UDI-Public00814548015392
Combination Product (Y/N)Y
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/20/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/27/2022
Device Catalogue Number400SMTXSFT0408
Device Lot NumberF81201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2018
Initial Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
Date Report Sent to FDA01/01/2005
-
-