MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE
|
Back to Search Results |
|
| Catalog Number |
3549216 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab received the device for evaluation on 12/8/2018.The analysis has begun, but is not complete at this time.When the investigation and analysis have been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed on 11/28/2018, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old female patient underwent breast augmentation revision with a mentor cpx 4 800cc breast tissue expander with suture tabs.The breast tissue expander was found to be leaking post procedure.As a result, the patient underwent unilateral removal and replacement with a similar mentor cpx 4 800cc breast tissue expander with suture tabs on (b)(6) 2018; catalog # 3549216, s/ns: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2019, mentor completed the investigation on the suspect medical device.Upon receipt by mentor, the device contained no fluid.White material was observed within the device and on the shell surface.During visual examination, mentor product analysis lab observed a rent on the posterior view of the device measuring approximately 0.8 cm in length.Microscopic examination of the rent edges gave no indications as to its cause.No other anomalies were found.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.Mentor product analysis lab concluded that the rent occurred sometime subsequent to the removal of the device from its protective packaging.The complaint of deflation was confirmed since a rent was found on the device.Nonetheless, a microscopic examination of the rent edges was performed, and it did not provide conclusive evidence of what could be the root cause.At this point it is not possible to determine the origin of the white material observed.There is no evidence that the issue is related with manufacturing.Because the risk of device deflation cannot be eliminated, mentor addresses this risk in the product insert data sheet (pids) by stating that these devices should not be considered lifetime devices; however, deflation trends for mentor saline-filled devices will continue to be monitored by quality assurance.Manufacturer¿s reference number: (b)(4).
|
| |
|