| Catalog Number |
2N8378 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing of three (3) clearlink sets split while in use on patients for ct (computerized tomography) dye administrations.The dyes were being injected at psi 4 (pounds per square inch).There was no report of patient injuries or medical interventions associated with these events.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not received for evaluation; therefore, a device analysis could not be completed.However, the customer reported that the set was used for a dye injection at 4 psi.The cause of the condition was determined to be a use error.Per the labeling for this device, the user is instructed not to connect any compressed gas device to intravenous injection sites.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|