| Model Number |
A34-34/C80-O20 V |
| Medical Device Problem Codes |
Leak/Splash (1354); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with a bifurcated stent graft and a suprarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).The three (3) month routine follow-up identified a type 1a endoleak and an indeterminate endoleak (possible type 3a or 2).A secondary procedure was completed.The physician elected to implant another suprarenal stent graft extension to resolve the reported the event.The patient was discharged.As of the date of this report, there have been no additional patient sequelae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.A clinical assessment of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Several attempts were made to obtain medical records and/or images but no response was received.As such, event determination, off label conditions, related patient harms and patient disposition could not be independently assessed.No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.Correction: conclusion code: remove code.Device code: remove code.
|
| |
|
Search Alerts/Recalls
|