During an evlt procedure, when the laser fiber was fired, it did not appear to be any steam bubbles.The device was withdrawn and it was noted the fiber appeared to be cracked.The device was set aside and a new of the same was used to complete the procedure.There was no harm or injury to the patient due to the event.It was reported the disposable device is available for return to the manufacturer for a device evaluation.
|
|
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of "fiber fractured" cannot be confirmed, without a sample evaluation.A root cause for the reported complaint description cannot be determined.A review of the angiodynamics' device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The directions for use, which is supplied to the end user with this catalog number, contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference complaint # (b)(4).
|