• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM LCP(TM) PLATE 8 HOLES/76MM; PLATE,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM LCP(TM) PLATE 8 HOLES/76MM; PLATE,FIXATION,BONE Back to Search Results
Model Number 247.372
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Impaired Healing (2378)
Date of Event 01/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Unknown date in 2018.(b)(6).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history: lot part: 247.372; lot: l770621; manufacturing site: (b)(4); release to warehouse date: 13.Feb.2018.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the patient underwent a hardware removal of a locking compression plate (lcp) plate, six (6) unknown cortex screws, and an unknown locking screw on a distal ulna fracture.This was due to non-union and the plate positioned incorrectly in the distal radial ulnar joint.The plate was removed, the edges of the fracture were corrected, and a new lcp plate was implanted.All hardware was removed successfully.The original surgery was an open fracture on (b)(6) 2018.It is unknown if there was a surgical delay.There was no patient consequence.This report is for one (1) 2.7mm lcp(tm) plate 8 holes/76mm.This is report 1 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM LCP(TM) PLATE 8 HOLES/76MM
Common Device Name
PLATE,FIXATION,BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8181532
Report Number2939274-2018-55485
Device Sequence Number11196223
Product Code HRS
UDI-Device Identifier10886982170873
UDI-Public(01)10886982170873
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K063049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number247.372
Device Catalogue Number247.372
Device Lot NumberL770621
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/28/2018
Initial Report FDA Received Date12/19/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age16 YR
Patient Weight68
-
-