MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE XPEDITION 48 EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE XPEDITION 48 EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1070250-48
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Intimal Dissection (1333)
Event Date 12/01/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of intimal dissection is listed in the xience xpedition, everolimus eluting coronary stent system (eecss), instructions for use (ifu) as a known patient effect of coronary stenting procedures.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event Description
It was reported the procedure was to treat a totally occluded lesion in the mid right coronary artery (rca).The lesion was pre0dilated with a semi complaint balloon and the 2.50x48mm xience xpedition stent was deployed at 16 atmospheres (atm).Post implantation, a distal edge dissection was noted.Another xience xpedition stent was used to treat the dissection.There was no clinically significant delay in the procedure and no adverse patient sequela.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION 48 EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key8181541
MDR Text Key130973540
Report Number2024168-2018-09887
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/24/2020
Device Catalogue Number1070250-48
Device Lot Number7100941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
-
-