| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-62-user had poor technique test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose test results.The customer is concerned with tests results from back-to-back blood tests of 60 and 422 mg/dl.Results were reported by the customer and could not be confirmed during review of meter memory results.The customer's expected fasting blood glucose test result range is 95 - 140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back-to-back blood test was not performed by the customer.The product is stored according to specification in the dining area.The test strip lot manufacturer's expiration date is 11/15/2020, and open vial date is two weeks.The meter memory was reviewed for previous test result history: (b)(6) customer called about the meter reading 60mg/dl then 2-5 minutes later reran her blood it read 422mg/dl as a fasting, i could not access the results in the memory.Customer was using alcohol prior to sticking her finger.
|
| |
|