| Model Number |
638RL32 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that this annuloplasty ring was implanted and explanted during the same procedure.While the physician was suturing the ring into the patient, the sutures "broke".The ring was successfully replaced with a mitral bioprosthetic valve.No additional adverse patient effects were reported. .
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received additional information that the sutures were broken when forming the valve for implant.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at medtronic¿s quality laboratory, visual examination revealed that the ring was discolored which indicates evidence of blood contact.There were tiny cuts and frayed wire observed.These observations did not contribute to the reported issue.Radiography revealed no fractures or ring deformation.Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Based on the returned product analysis and received information.Without return of the ring holder, a thorough analysis of the broken suture could not be performed.However, there is no indication that the event was due to manufacturing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|