| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(bathroom).Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low and erratic blood glucose test results.The customer is concerned with tests results of 58 and 61mg/dl.The expected fasting blood glucose test result range is 70-80mg/dl.The customer did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the bathroom.Test strip lot manufacturer's expiration date is 04/23/2019, and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history:
(b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = t) internal report (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Correction:.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low and erratic blood glucose test results.The customer is concerned with tests results of 58 and 61mg/dl.The expected fasting blood glucose test result range is 70-80mg/dl.The customer did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the bathroom.Test strip lot manufacturer's expiration date is 04/23/2019 and open vial date is 10/30/2018.The meter memory was reviewed for previous test result history: result 1:123mg/dl date:(b)(6) time:1:53pm fasting, result 2:61mg/dl date:(b)(6) time:1:51pm fasting, result 3:108mg/dl date:(b)(6) time:2:00am fasting, result 4:126mg/dl date:(b)(6) time:10:17pm unknown, result 5:58mg/dl date:(b)(6) time:1:53pm fasting.
|
| |
|