ENCORE MEDICAL L.P. 3D KNEE; TIBIA, STEMMED, POROUS SIZE8 L 3D Back to Search Results
Model Number 391-09-608
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/05/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a loose size 8 3d porous tibia and size 8 3d femur.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to the loosening.The previous surgery and the revision detailed in this investigation occurred 9 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was a significant adverse event to the patient.There was a 3 minute delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: poor bone density, inadequate soft tissue support, degenerative tissue, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3D KNEE
Common Device Name
TIBIA, STEMMED, POROUS SIZE8 L 3D
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8182231
Report Number1644408-2018-01165
Device Sequence Number19207966
Product Code JWH
UDI-Device Identifier00888912118897
UDI-Public(01)00888912118897
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K020114
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2014
Device Model Number391-09-608
Device Catalogue Number335-01-108
Device Lot Number53911206
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/05/2018
Supplement Date Received by Manufacturer01/10/2019
Initial Report FDA Received Date12/19/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
236-01-108, LOT 53935709; 236-01-108, LOT 53935709; 335-01-108, LOT 53911206; 391-09-608, LOT 54002571
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age72 YR
-
-