• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6531
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/07/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon detachment occurred.A 12mm x 4.00mm nc quantum apex balloon catheter was selected for used to dilate the lesion; however, it was noted that the balloon ripped off.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the inflation lumen is separated, and the distal section of the device is missing.The missing parts are the distal inflation lumen, entire guidewire lumen, balloon, marker band, and tip.There are multiple kinks along the hypotube and on the core wire.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the distal section of the device is missing.
 
Event or Problem Description
It was reported that balloon detachment occurred.A 12mm x 4.00mm nc quantum apex balloon catheter was selected for used to dilate the lesion; however, it was noted that the balloon ripped off.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8182292
Report Number2134265-2018-64513
Device Sequence Number13850004
Product Code LOX
UDI-Device Identifier08714729783305
UDI-Public08714729783305
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2021
Device Model Number6531
Device Catalogue Number6531
Device Lot Number0022788757
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Initial Date Received by Manufacturer 12/11/2018
Supplement Date Received by Manufacturer01/31/2019
Initial Report FDA Received Date12/19/2018
Supplement Report FDA Received Date02/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-