| Model Number |
6531 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon detachment occurred.A 12mm x 4.00mm nc quantum apex balloon catheter was selected for used to dilate the lesion; however, it was noted that the balloon ripped off.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the inflation lumen is separated, and the distal section of the device is missing.The missing parts are the distal inflation lumen, entire guidewire lumen, balloon, marker band, and tip.There are multiple kinks along the hypotube and on the core wire.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the distal section of the device is missing.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon detachment occurred.A 12mm x 4.00mm nc quantum apex balloon catheter was selected for used to dilate the lesion; however, it was noted that the balloon ripped off.No patient complications were reported.
|
| |
|