Device evaluated by mfr.: returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the inflation lumen is separated, and the distal section of the device is missing.The missing parts are the distal inflation lumen, entire guidewire lumen, balloon, marker band, and tip.There are multiple kinks along the hypotube and on the core wire.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the distal section of the device is missing.
|