ENCORE MEDICAL L.P. EMPOWR KNEE; EMPOWR 3D KNEETM INS, 6L 16MM, VE Back to Search Results
Model Number 341-16-706
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 12/05/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient complaining of pain and discomfort.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to pain and discomfort.The previous surgery and the revision detailed in this investigation occurred 16 days apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain and discomfort.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: degenerative bone disease, patient bone deterioration, ageing of the patient, patient activities or trauma.Due to the short time between the previous and revision surgery, it is also possible that the event may have occurred due to improper implant selection or patient non-compliance with medical instructions.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR KNEE
Common Device Name
EMPOWR 3D KNEETM INS, 6L 16MM, VE
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8182320
Report Number1644408-2018-01174
Device Sequence Number1225476
Product Code JWH
UDI-Device Identifier00888912167130
UDI-Public(01)00888912167130
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model Number341-16-706
Device Catalogue Number341-16-706
Device Lot Number150T1002
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/05/2018
Supplement Date Received by Manufacturer12/28/2018
Initial Report FDA Received Date12/19/2018
Supplement Report FDA Received Date01/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
-
-