• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING SL; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING SL; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24702
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.
 
Event or Problem Description
It was reported that balloon rupture occurred.During a left popliteal artery dilatation, a 3.0mm x 100mm x 150cm sterling sl balloon catheter was advanced.However, on initial inflation at 10 atmospheres, the balloon ruptured.The device was completely removed from the patient and the procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling sl balloon catheter.The balloon was loose and there was blood and dried contrast in the inflation lumen (shaft).The tip, balloon, markerbands, proximal weld, and inner/outer shaft were microscopically and visually inspected.Inspection revealed a burst in the balloon material that was 11 mm in length.Inspection of the remainder of the device found no other damage or irregularities.
 
Event or Problem Description
It was reported that balloon rupture occurred.During a left popliteal artery dilatation, a 3.0mm x 100mm x 150cm sterling sl balloon catheter was advanced.However, on initial inflation at 10 atmospheres, the balloon ruptured.The device was completely removed from the patient and the procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING SL
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8182374
Report Number2134265-2018-64316
Device Sequence Number1587923
Product Code DQY
UDI-Device Identifier08714729782292
UDI-Public08714729782292
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K093720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2020
Device Model Number24702
Device Catalogue Number24702
Device Lot Number0022464275
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Initial Date Received by Manufacturer 12/10/2018
Supplement Date Received by Manufacturer01/17/2019
Initial Report FDA Received Date12/19/2018
Supplement Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
-
-