• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNC Back to Search Results
Catalog Number 625-0T-32E
Medical Device Problem Codes Degraded (1153); Unstable (1667); Malposition of Device (2616)
Health Effect - Clinical Codes Injury (2348); Ambulation or Postural Difficulties (2544)
Date of Event 11/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding instability and wear involving trident liner was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned medical records received and evaluation: no medical records were received for review with a clinical consultant device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.The following devices were also listed in this report: trident psl ha cluster 50mm; cat# 542-11-50e, lot# 21866401.6.5 cancellous bone screw 30mm, cat# 2030-6530-1 ; lot# 7mmmdd.6.5 cancellous bone screw 30mm, cat# 2030-6530-1 ; lot# 7mmmdd.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
Patient had complained of subluxation and grinding of hip joint that was replaced in 2007.When surgeon operated and dislocated this joint the ceramic liner was found to have a worn superior aspect of the liner.Surgeon performed an acetabular revision.The worn liner was noticed intraoperatively, however the patient had been complaining of instability previously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT ALUMINA INSERT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNC
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8182673
Report Number0002249697-2018-04085
Device Sequence Number16721206
Product Code MRA
UDI-Device Identifier04546540516848
UDI-Public04546540516848
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2012
Device Catalogue Number625-0T-32E
Device Lot Number22272301
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2018
Initial Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
-
-