| Catalog Number |
49350020 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.Device disposiiton is unknown.
|
| |
|
Event or Problem Description
|
|
This pi is for the second revision of patient's foot arch, and hexapod strut issue.As reported: ".We replace that foot arch with a new one and he[patient] breaks that one the following week.Bring him back to office and revise.While revising we notice the pin in the hexapod strut was starting to come out".Rep indicated patient factor: diabetic.Revision, broken foot arch, hexapod strut malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that hexapod strut, short hoffmann lrf 105mm-139mm was alleged of 'implant breakage after surgery' could be confirmed.Based on investigation, the root cause was attributed to be manufacturing and supplier related.The failure was caused by manufacturing components out of specifications.The event involved a product problem indicating a non-conformity which is investigated and addressed through a nc and a capa.No further action is required at this time.Product surveillance will continue to monitor complaints of this type for adverse trends.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
This pi is for the second revision of patient's foot arch, and hexapod strut issue.As reported: "we replace that foot arch with a new one and he[patient] breaks that one the following week.Bring him back to office and revise.While revising we notice the pin in the hexapod strut was starting to come out".Rep indicated patient factor: diabetic.Revision, broken foot arch, hexapod strut malfunction.
|
| |
|
Search Alerts/Recalls
|