| Catalog Number |
4350 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an unknown procedure on (b)(6) 2018 and the absorbable adhesive barrier was used.The device was found unsealed for the sterile packaging, and no sealing glue.There was no patient consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The following additional information was received: during the surgery, 1 unit was found unsealed.Two unsealed units were found after surgery when checked their stock, which are from the same lot.Malfunctions reported via manufacturer report # 2210968-2018-77919, 2210968-2018-77922, 2210968-2018-77923.
|
| |
|
Additional Manufacturer Narrative
|
|
Date sent to the fda: 03/04/2019.Additional summary: one (1) opened sample of product was returned for analysis.Sample was numbered to evaluate the variety of failure modes as different pouch conditions were found.Sample # 2 was found closed on the chevron side but was unsealed on the non-chevron side of the sterile packaging.It can be concluded that this was tone of the units found in the inventory after the subject unit was found as it has an open seal and it was closed in the chevron side.
|
| |
|