Model Number MMT-751NAP
Medical Device Problem Codes No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Moisture or Humidity Problem (2986)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was unknown.The customer reported that insulin pump was exposed to moisture.The customer was using an old insulin pump that she got the button error on.The customer was assisted with troubleshooting for fluid exposure.The customer reported that the buttons stopped working the next day and the screen went blank.The customer was advised to discontinue use of the insulin pump and to revert to the backup plan.The customer was advised the pump will be replaced as per troubleshoot.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
All buttons functioning properly.However, found slightly corroded keypad traces.No button error alarm during testing.Insulin pump passed displacement test, self test, off no power test and all operating currents tested within the spec range.Insulin pump was monitored and tested with no blank display noted.Insulin pump received with moisture damage on electronics assembly noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8184178
Report Number2032227-2018-77735
Device Sequence Number2141953
Product Code OZO
UDI-Device Identifier00643169736092
UDI-Public(01)00643169736092
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAP
Device Catalogue NumberMMT-751NAP
Device Lot NumberA6751NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Initial Date Received by Manufacturer 12/06/2018
Supplement Date Received by Manufacturer02/06/2019
Initial Report FDA Received Date12/20/2018
Supplement Report FDA Received Date02/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR