(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Stents: 3 unknown xience sierra stents, 3.25 x 28 mm xience sierra stent.(b)(4).The additional 3.25 x 28 xience sierra device referenced is being filed under a separate medwatch report.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The xience sierra everolimus eluting coronary stent system (eecss), domestic, instructions for use states: when treating multiple lesions within the same vessel, stent the distal lesion prior to stenting the proximal lesion.Stenting in this order obviates the need to cross the proximal stent during placement of the distal stent, and reduces the chances of damaging or dislodging the proximal stent.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the stent interacted with the previously deployed stent while attempting to cross/position (against instructions for use) causing the reported failure to advance, device entrapment and subsequent stent dislodgement with the patient effects of device embedded and additional therapy/non-surgical treatment.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
It was reported that the procedure was to treat a moderately tortuous right coronary artery.A 3.25 x 28 mm xience sierra stent was implanted along with 3 other unspecified xience sierra stents without any device issues.As a gap was noted between 2 stents, a 2.25 x 15 mm xience sierra stent was advanced to the lesion; however, the stent became caught with the 3.25 x 28 mm sierra stent.It was noted under angiography that the 3.25 x 28 mm sierra stent was not damaged.The 2.25 x 15 mm sierra stent delivery system (sds) was then moved back and forth to remove the stent, and it dislodged.Therefore, an unspecified trek and a 1.5 x 12 mm mini trek balloon catheter were used to crush the stent to the 3.25 x 28 mm stent.Post-dilatation was then performed with a 3.5 x 15 mm nc trek balloon catheter to successfully complete the procedure.No additional information was provided.
|