| Model Number |
1365-06-100 |
| Medical Device Problem Codes |
Corroded (1131); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Edema (1820); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Synovitis (2094); No Code Available (3191)
|
| Date of Event |
11/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
| |
|
Event or Problem Description
|
|
The patient was revised to address pain.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Udi: (b)(4).No code available use for medical device removal.
|
| |
|
Event or Problem Description
|
|
After review of medical records patient was revised to addressed left hip component wear and lateral soft tissue mass from adverse soft tissue reaction.Operative notes indicated soft tissue swelling.There was a significant synovitis with metallosis brown staining of the articular portion of the soft tissues.Corrosion and wear within the trunnion junction and articulation of the head to the stem was noted.Doi: (b)(6) 2010; dor: (b)(6) 2018; left hip.
|
| |
|