MAUDE Adverse Event Report: ST. JUDE MEDICAL SUPREME¿ ELECTROPHYSIOLOGY CATHETER, QUADRIPOLAR, 120 CM LENGTH, CRD¿ CURVE, ELE; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SUPREME¿ ELECTROPHYSIOLOGY CATHETER, QUADRIPOLAR, 120 CM LENGTH, CRD¿ CURVE, ELE; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 401449
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2018
Type of Reportable Event Malfunction
Event or Problem Description
During a supraventricular tachycardia procedure, the catheter shaft fractured, exposing the internal components after being maneuvered.The device was replaced and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
One 6f, quadripolar, crd, supreme ep catheter was received for evaluation.The reported shaft fracture was confirmed.The shaft of the catheter was noted to be separated between the grey and black braided shaft portion due to an inefficient bonding technique.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPREME¿ ELECTROPHYSIOLOGY CATHETER, QUADRIPOLAR, 120 CM LENGTH, CRD¿ CURVE, ELE
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key8184539
Report Number3008452825-2018-00442
Device Sequence Number14139398
Product Code DRF
UDI-Device Identifier05414734201414
UDI-Public05414734201414
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K894500
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number401449
Device Lot Number6476954
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer04/05/2019
Initial Report FDA Received Date12/20/2018
Supplement Report FDA Received Date04/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-