TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.Customer declined replacement product at this time.Most likely underlying root cause: mlc-62-user had poor technique test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the initial concern was resolved - unable to establish contact with the customer at this time.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results of 305 and 319 mg/dl.The customer's expected fasting blood glucose test result range is 150 - 175 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back-to-back blood test was performed by the customer non-fasting and produced test results of 240 mg/dl and 243 mg/dl using truetrack meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 09/15/2020, and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6) customer states the meter is reading high blood results and only concerned with this.Customer states their normal fasting range is 150mg/dl-175mg/dl fasting.Customer adjusted meter's position from horizontal position while testing to a vertical position.Customer states they adjusted their testing technique by rubbing hands below waist level prior to pricking finger.Customer denies the need for medical attention at the time of call and denies signs and symptoms of hypoglycemia or hyperglycemia.Customer declined replacements and will implement new testing techniques and i will follow back-up with customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8184731
Report Number1000113657-2018-01272
Device Sequence Number2879388
Product Code NBW
UDI-Device Identifier00021292001018
UDI-Public(01)00021292001018
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/15/2020
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRV5257
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2018
Initial Report FDA Received Date12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-